Canada
& USA Toll Free
1 800 279-9912
|
|
|
|
Temperature
Monitoring System 12-Channel Scanning
- Store and recall up
to 4680 sets of readings with time and date
- Automatically or manually
display, record and output temperature readings
- Field calibrate each
probe separately - achieve accuracies of ± 0.1%
of reading!
- Bidirectional RS-232
for setup or PC datalogging
- Includes a real-time
clock, nonvolatile memory, and parallel printer output
(benchtop only)
- Immune to RF interference
Our Temperature Monitoring
System is ideal when simultaneous monitoring of temperatures
during different process steps or phases is critical,
these new 12-channel scanning thermometers.
The wall mount Temperature Monitoring System is ideal
for permanent installation and continuous monitoring of
your processes.
The benchtop Temperature Monitoring System accepts most
any power source giving you great flexibility as to where
and when it may be used. Both models function as a datalogger
- store data in the non-volatile memory and later download
to your printer or PC. Link both models in series to expand
the number of channels you can scan. Each of the 12 thermocouple
inputs are programmable to scan from every four seconds
to hourly.
BENCHTOP THERMOMETER
Parallel printer connection
allows you to connect the thermometer to a printer - make
a hard copy of your data as a permanent record
Power unit with included
115 VAC or 230 VAC adapter or any 10 to 25 VDC or 7 to
20 VAC 500 mA power supply. (user supplied)
Accepts thermocouple probes
with miniconnectors for quick and easy probe changes.
WALL-MOUNT
THERMOMETER
Use high and low relays
for on/off control; relays are rated at 250 VAC 2A or
2A at 30 VDC resistive.
Features an easy-to-view
drip proof housing with all wire connections on bottom.
Power the unit with 100-240
VAC, 50/60Hz
INSTRUMENT
FEATURES for both models
- Scan and log up to 12
thermocouple inputs from any combination of J, K, T,
E, N, B, R, and S thermocouple probes.
- Thermocouple type is
individually selectable per channel.
- Datalogging feature
automatically stores up to 4680 sets of readings.
- Download readings or
stored data to your computer.
- Link up to 8 units together
to expand channel capacity.
- Sequentially captures
temperature data, allowing you to monitor and record
process variations - ideal for statistical process control.
- Easy-to-use front panel
MENU key lets you program datalogging and print intervals,
scan rate, date and time, temperature scale (°F,
°C, K, °R), resolution (0.1 or 1°), channels
displayed, thermocouple type, and data output parameters.
- Each channel features
HOLD, STORE, MAX and MIN (measures maximum and minimum
values), RECALL and AVERAGE buttons.
- Calibrate each channel
separately at either one or two points to minimize probe
offset. Parameter setup and calibration offsets are
stored in nonvolatile memory.
- Each scanner is factory
calibrated to NIST-traceable standards.
MULTIPLE SAFETY
FEATURES for both models
- MENU key lets you select
high or low alarm set points for each channel. Easily
enable or disable audible alarm and set alarm hysteresis;
program for manual or automatic alarm reset.
- Thermocouple inputs
are electrically isolated from each other and from data
outputs - provides safe and accurate measurements even
when probes are exposed to various electrical conditions.
PHYSICAL FEATURES
for both models
- Large 3/4" high, 14-segment,
12-character alphanumeric LCD indicates channel number
or name, temperature reading, and temperature units.
- Indicator light warns
you of open probes or out-of-range readings.
- Benchtop models accept
thermocouple probes with ANSI miniconnectors for quick
hookup and changes.
- Wall-mount models accept
thermocouple probes with stripped ends.
COMPUTER/PRINTER
CAPABILITIES for both models
- Use printing function
to send current readings directly to your computer and
printer at chosen intervals.
- Vary datalogging and
printing intervals from 4 seconds to 99 minutes, 59
seconds.
- Features RS-232 (for
both units), and parallel printer output of date and
time (benchtop only), channel number and name, and temperature
readings in a tabular format.
- Includes RS-232 cable.
Parallel printer cable is not included (sold separately)
Remember to order the
proper probes for your application. Barnant offers a wide
variety of thermocouple probes (J, K, T, E, N, R, S, and
B) wire, connectors and other accessories to insure your
temperature scanner will provide the ultimate of function
and performance.
Temperature
Monitoring System Specifications:
Ranges:
J: -200 to 1000°C (-392 to
1832°F)
K: -250 to 1372°C (-418 to
2501°F)
T: -250 to 400°C (-418 to 752°F)
E: -250 to 1000°C (-418 to
1832°F)
N: -250 to 1300°C (-418 to
2372°F)
R: 0 to1768°C (32 to 3214°F)
S: 0 to 1768°C (32 to 3214°F)
B: 200 to 1800°C (392 to 3272°F)
Resolution: 0.1°/1° selectable up to
999.9°, 1° above 1000°; autorange to 1°
below -150°C
Accuracy: J, K, T, E, N: ±0.1% of rdg.,
±0.5°C (±0.8°F) above -150°C;
±0.25% of rdg., ±1°C (±2°F)
below -150°C
R, S, B: ±0.1% of reading, ±2°C (±4°F)
Display: 12 character alphanumeric LCD
Display update rate: 3 sec/channel
Scan rate: from 4 sec/12 channels to 99 min
59 sec/12 channels
Alarm output: (wall-mount units only) SPDT
high/low relays are rated at 250 VAC 2A or 2A at 30
VDC resistive.
Digital output: Both models: RS-232, ASCII
format, baud rate 19.2 K, RJ11connector;
Benchtop unit only: Centronics parallel printer output;
DB25F connector
Dimensions: Benchtop: 7 1/2"L x 10 1/2"W x
3 1/2"H
Wall-mount: 5 1/2"L x 11 1/4"W x 8 1/2"H
Shpg wt: Benchtop: 3 lb (1.4 kg)
Wall-mount: 4 lb (1.8 kg)
Power: Wall-mount: 90-240 VAC, 50/60Hz
Bench-top: 10 to 25 VDC, or 7 to 20 VAC 500mA. (optional,
user supplied)
|
ORDERING INFORMATION
Cat
No |
Description |
Include
Power Supply |
692-0000
692-0230 |
Benchtop |
115
VAC 50/60 Hz
230 VAC 50/60 Hz |
692-1000
692-1230 |
Wall-mount |
115
VAC 50/60 Hz
230 VAC 50/60 Hz |
For
Quick Delivery
From Stock Call
CANADA